Stakpure OmniaPure UV/UF Ultrapure Water System, 2 l/min, Molecular & Microbiology, Cell Culture Media

No.
PE-0005
Copy SKU
Note: Illustrations may differ or contain accessories.
Manufacturer

Delivery time

2-3 weeks

VE

1 Stück

Price on request
plus VAT & Shipping
Did you know? All immediately available items ordered by 10:30 a.m. leave the warehouse on the same day.
The OmniaPure ultrapure water system is particularly recommended for life science and analytical applications where ultrapure water of the highest quality is required.
With an integrated pretreatment unit, the system ensures reliable analysis results and thus reduces follow-up costs. The ultrapure water (category ASTM I) is withdrawn by a flexible dispenser, with digital control.
The model OmniaPure UV_UF system reduces the TOC value by an integrated UV lamp.
In addition, it can remove endotoxins, RNase and DNase through its integrated UV lamp with downstream ultrafiltration.
The ultrapure water systems require pre-purified water and cannot be connected directly to the drinking water line.
Scope of supply: ultrapure water system including pretreatment cartridge, ultrapure water cartridge & sterile filter capsule.

Typical application:
AAS, IC, ICP, buffer and media preparation

Specifications:
ASTM I Conductivity: 0.055 | 0.055 | 0.055 µS/cm Resistivity: 18.2 | 18.2 | 18.2 MΩ*cm TOC value*: 5 - 10 | 1 - 5 | 1 - 5 ppb Sampling capacity: up to 2 | up to 2 | up to 1.6 l/min Programmable volume dosing: 0.01 - 60 | 0.01 - 60 | 0.01 - 60 Endotoxins*: - | - | 0.001 EU/ml RNase*: - | - | 4 ng/ml DNase*: - | - | 0.01 pg/µl Bacterial /Particle: < 1 | < 1 cfu/ml (with final filter)
*depending on feed water quality
Druckansicht (PDF) Druckansicht (PDF)
No. PE-0005
Manufacturer (Brand) Stakpure
old neoLab article no. 2-1230
EAN 4066292549883
Transport temperature Room temperature
autoclavable at 121°C No
sterile No
Display (reading, setting) digital
Operation via Membrane keypad
for ambient temperature 2 to 35 °C
Application analytics Molecular and Microbiology|Cell Culture Media
Application product water General laboratory applications (e.g. autoclaving. media preparation. etc.)|Analytics|Life Sciences
Bacteria < 1 cfu/ml
DNase 0.01 pg/µl
Feed water pressure 0.1 to 6 bar
Endotoxins/ Pyrogens 0.001 EU/ml
Withdrawal possibilities product water type I with dispenser
Spare part PE-0043|PE-0044|PE-0041
Marking by manufacturer with ASTM
Conductivity product water type I 0.055 µS/cm
Model High purity/clean water system stakpure OmniaPure UV/ UF
Production capacity water type I 1.6 l/min
Quality product water Type I (ultrapure water)
Feed water quality pre-purified water
RNase 4 ng/ml
TOC 1 to 5
Resistance product water type I 18.2 MΩ*cm